A calibration curve links an instrument's signal to concentration using standards. Enter the concentrations and signals and this calculator fits the best straight line, gives R squared and finds the concentration of an unknown.

Formula

Signal = slope × concentration + intercept, so concentration = (signal − intercept) ÷ slope

  • The line is fitted by least squares
  • R² near 1 means a good straight line

Standards 0, 2, 4, 6, 8, 10 with signals 0.01, 0.21, 0.40, 0.62, 0.80, 1.01 give a slope of 0.0999, an intercept of 0.0090 and R² = 0.9996. An unknown signal of 0.50 is 4.92

Calibration Curve Calculator

Least-squares line through the standards, y = slope × concentration + intercept. Keep the unknown inside the range of the standards.

Stay Inside the Range

The line is only trustworthy between your lowest and highest standard. Outside that, the response may curve and the result is an extrapolation. If an unknown is too strong, dilute it and measure again.

Check the Fit

R² close to 1 is necessary but does not prove the line is right. Plot the points, and look for a curve or an outlier. Use at least five standards spread across the range, and include a blank.

Frequently Asked Questions

How do I make a calibration curve?
Measure the signal of several standards of known concentration and fit a straight line through them.
What does R² tell me?
How closely the points follow the line. Values near 1 mean a good fit.
Can I use a signal outside the standards?
It is risky, because the response may no longer be linear.